Sysmex XE-2100 Performance Evaluation
Performance Evaluation of the Sysmex XE-2100 Hematology Analyzer
JERELYN WALTERS,1
PATRICIA GARRITY 2
1 ACL Laboratories, West Allis;
2 Milwaukee Area Technical College, Milwaukee, Wisconsin
A new hematology analyzer, the XE-2100 (Sysmex, Kobe, Japan), was evaluated for performance in the core laboratory of a multiple-hospital system. Complete blood count performance was evaluated in comparison to the SE-9000 (Sysmex). Differential
performance was evaluated according to the National Committee for Clinical Laboratory Standards (NCCLS) document H20-A 1. Precision, carryover, correlation, and linearity studies all showed excellent results. Stability studies showed no significant change in differential parameters for a period of 68 hours at both refrigerated and room temperatures. It appears that the stability of the automated differential on the XE-2100 now exceeds the stability of red
blood cell size parameters that were stable for 36 hours when refrigerated. Results comparing the XE-2100 automated differential with the 400-cell manual differential were very good. The new nucleated red blood cell (NRBC) measurement showed excellent precision and linearity. Correlation with manually enumerated NRBCs exceeded acceptable statistical standards. The evaluation led to the conclusion that the XE-2100 is an accurate and precise high-speed analyzer. Lab Hematol. 2000;6:83-92.
LH.6.2.5.Walters.pdf
Copyright 1995-2010 - Carden Jennings Publishing Co., Ltd. All rights reserved. The material available at this site is for educational purposes only and is NOT intended for any diagnostic, clinically related, or other purpose. Carden Jennings Publishing Co., Ltd., assumes no responsibility for any use or misuse of this material and makes no warranty or representation of any kind with respect to the material available at this site.
|
|